Toxcity Chamber
Model PTE005
Toxicity Chamber consists of an airtight enclosure of about
0.7m³ volume lined with opaque plastic sheeting having a hinged, fitted with a
transparent plastic panel. Some gaseous products of combustion may react with
or be absorbed on the walls of the chamber. So the inner wall has the lining of
polypropylene and with polycarbonate transparency window is provided. The
chamber is fitted with a vacuum pump forced air extraction system, which can be
closed at the exit from the chamber when required. The chamber is fitted with
sampling positions such that the air tightness of the chamber is not impaired.
The chamber is containing a mixing axial fan inside the chamber at horizontally
and centrally inside the chamber at roof level to ensure rapid mixing of
combustion products capable of being switched on and off externally. Burner is
to be capable of achieving a flame approximately 100 mm in height. The burner
is fitted at the center of the chamber. There is a provision is to be made for
igniting and extinguishing the burner from outside the chamber. A digital
temperature indicator is also mounted for the measurement of temperature of the
flame of the burner. The specimen support is an annulus cut from a sheet steel
nominal thickness 2 mm of approximately 100 mm overall diameter with a 75 mm
diameter hole carrying temperature resistance wires approximately 10 mm apart
to form a lattice.
The complete assembly is equipped with a non-combustible
side support arm to give a tennis racket appearance. This sample holder
assembly has the provision of sliding of sample without the leakage of the
chamber. There are two flow meters are fitted to the instrument to fine control
of the supply of natural gas and air. All the parts are either polished or
painted to give the corrosion resistance long life. Gas detection tubes and its
pumps are optional which will be supplied at separate order as per customer
requirement.